H3: product analysis: evaluation could not reproduce the reported malfunction as the temperature was within specification at 87°f.The likely cause was identified as worn bearings.It was also noted that the laser markings faded, housing scratched, gearing's are dry and had insufficient lubrication.B3: date of notification: 2024-05-14 (date of event or estimated date of incident not provided to manufacturer).This regulatory report is being submitted due to a retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|