H3: product analysis: evaluation was not able to preform the reported malfunction of overheating due to motor is not working gives error code 15.It was also noted that cosmetic condition of product is not okay, cable has nicks and kinks and connector is dented.It was also noted that hi-pot test failed, depth test failed, corrosion and motor failure.This regulatory report is being submitted due to a retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|