• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRAUB MEDICAL AG ROTAREX; THROMBECTOMY & ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRAUB MEDICAL AG ROTAREX; THROMBECTOMY & ATHERECTOMY Back to Search Results
Catalog Number 80201
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H11: section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a recanalization procedure in the left external iliac artery via contralateral approach, the tip of the device allegedly broken.It was further reported that the device was allegedly detached.Reportedly, the detached part could be snared.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: no physical sample / images / videos were received.A physical investigation was not possible.Due to no sample / images / videos received, the reported malfunction cannot be confirmed.A clear root cause could not be established.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a recanalization procedure in the left external iliac artery via contralateral approach, the tip of the device allegedly broken.It was further reported that the device was allegedly detached.Reportedly, the detached part could be snared.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the catheter was returned for evaluation, and thus physical investigation was performed for the catheter.During physical investigation the helix was found broken at 25 cm distance from the tip of the catheter.The tube of the catheter had a strong kink and a hole at the same position at 25 cm from the tip of the catheter.The broken part of the helix and the head part were missing.It was not possible to specify the root cause further using the information provided by the user.It can be confirmed that the helix was broken.A clear root cause could not be established.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a recanalization procedure in the left external iliac artery via contralateral approach, the tip of the device allegedly broken.It was further reported that the device was allegedly detached.Reportedly, the detached part could be snared.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Common Device Name
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
STRAUB MEDICAL AG
straubstrasse 12
ch-7323
wangs SG 73 23
SZ  SG 7323
Manufacturer (Section G)
STRAUB MEDICAL AG
straubstrasse 12
ch-7323
wangs SG 73 23
SZ   SG 7323
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key21686084
Report Number3008439199-2025-00050
Device Sequence Number18336492
Product Code MCW
UDI-Device Identifier07640142810292
UDI-Public(01)07640142810292
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K211738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number80201
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2025
Supplement Date Received by Manufacturer04/23/2025
05/09/2025
Initial Report FDA Received Date03/25/2025
Supplement Report FDA Received Date05/08/2025
05/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-