Catalog Number 257000000
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that the both broach handles were not holding the broaches as they should be.Extremely loose fit.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device revealed that the summit universal broach handle had signs of extensive and repeated use over the years.It is worth noting that the assembly section was worn out.A functional test was conducted using a laboratory sample mating device, and the summit universal broach handle was successfully assembled, however, looseness was observed in the assembly preventing the device from securely holding its mating device.Considering the device's age (18+ years), the observed condition appears to be consistent with the effects of repeated use and servicing over the years.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the summit universal broach handle would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (ak0606) provided is not a valid finished goods lot number.H11 additional narrative: corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the both broach handles were not holding the broaches as they should be.It was an extremely loose fit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT UNIVERSAL BROACH HANDLE
Common Device Name
HIP INSTRUMENTS : HANDLES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21686214
Report Number1818910-2025-04473
Device Sequence Number18355189
Product Code HTQ
UDI-Device Identifier10603295142843
UDI-Public(01)10603295142843(10)AK0606
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number257000000
Device Lot NumberAK0606
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/27/2025
Supplement Date Received by Manufacturer03/26/2025
Initial Report FDA Received Date03/25/2025
Supplement Report FDA Received Date03/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/15/2006
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
-
-