• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED LARGO ERGO; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED LARGO ERGO; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number D4240
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
The sales representative reported on behalf of the customer that the d4240, ergo 2-button shaver handpiece was found during pre-op testing on approximately 14mar25 when it was reported ¿gets hot and stops running.¿.There was no report of injury, medical intervention, or extended hospitalization for the patient or user.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor per regulatory requirements to help ensure patient safety.
 
Additional Manufacturer Narrative
An evaluation of the returned device found motor ¿ hall cell failure and gears ¿ seized / binding / grinding.Preventative maintenance is not overdue.Repair/evaluation completed.Final test completed and met all specifications.Root cause cannot be determined, however, based upon evaluation of the device; a possible cause of this event could be the gears seizing / binding / grinding.The service history was reviewed, and no prior data was found.A device history record (dhr) review found no abnormalities that would contribute to this reported event.A two-year review of complaint history revealed there has been a total of (b)(4) complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised to continually check handpiece for overheating.If overheating is sensed, immediately discontinue use and return equipment for service.Overheating of the blade or bur may cause damage to the blade or bur and may cause burns or thermal necrosis.Running the shaver blade or bur without fluid flow (dry) may cause damage to the handpiece or may cause the bur hubs to melt due to excessive heat.Do not apply excessive loading on the shaver blade or bur.Cutting performance is not increased with force.Excessive force or using shaver blades or burs as a lever can cause damage to the device including permanent deformation, shedding of metal (wear), motor seizure, and overheating.Run the handpiece for less than 5 seconds at a speed less than 1500 rpm to observe any abnormal noises, vibrations or heat rise.We will continue to monitor per regulatory requirements to help ensure patient safety.
 
Event or Problem Description
The sales representative reported on behalf of the customer that the d4240, ergo 2-button shaver handpiece was found during pre-op testing on approximately (b)(6) 2025 when it was reported ¿gets hot and stops running.¿.There was no report of injury, medical intervention, or extended hospitalization for the patient or user.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ERGO
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
CONMED LARGO
11311 concept blvd
largo FL 33773
Manufacturer (Section G)
CONMED LARGO
11311 concept blvd
largo FL 33773
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key21686352
Report Number1017294-2025-00035
Device Sequence Number9420130
Product Code GEY
UDI-Device Identifier10845854013144
UDI-Public(01)10845854013144(11)241202
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K050519
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD4240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2025
Supplement Date Received by Manufacturer04/25/2025
Initial Report FDA Received Date03/25/2025
Supplement Report FDA Received Date05/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2024
Is the Device Labeled for Single Use? (Y/N) No
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-