This report is being submitted to relay additional information and device evaluation.Zimvie received the reported implant for evaluation.Visual evaluation was performed, the implant had signs of use/wear with debris on its external threads.The implant had a crown attached.There was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was inadequate treatment planning.Therefore, based on the available information, device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable with all the available information provided.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|