It as reported that during bipolar resection of the prostate (tur prostate), the patient suddenly received electric shocks when the current was applied, which caused muscle twitching.At the same time, there were screen failures.The resection snare was replaced immediately.Damage to the insulation of the snare was noticed.A short time later, the same problem occurred with the first replacement snare.Here too, the insulation on the sling was missing.When using the 2nd replacement sling, it was observed that the insulation of the sling was pulled inwards as it was extended and tightened until it had completely slipped out of place.The operation was then aborted.The user reported a clinically relevant increase in operating time of approx.45 minutes.In addition, the medical purpose of the application could not be fulfilled as the prostate could not be completely resected.
|
|
The cutting electrode bipo 24fr 12/30, part id: 46221313, batch# 21005386 was sent to richard wolf (rwgmbh) for evaluation.During examination, the test series have shown that the fault pattern described in the incident can only occur when a tensile load is applied in the direction of the distal loop.This is only conceivable when removing the electrode from the working element if, for example, the user does not simultaneously activate the locking mechanism on the lock body.In this regard, the user is advised in the instructions for use, ga-d342 / en / us / v11.0 / 2023-08, the applicable warnings for these errors.Chapter 7: use - contains warning and caution regarding the device limited strength.Exerting excessive force will cause damage, impair the function, and therefore endanger the patient.Immediately before and after each use, check the products for damage, loose parts, and completeness.Chapter 7.1.9 describes in detail how to insert the electrode into the working element, locking, and releasing.It also includes an illustration of how to hold the electrode with your fingers.Incorrect installation by the user - in the sense of a potential application error - was taken into account in risk-368 in the context of electromagnetic energy.The risk of insufficient insulation due to improper installation is addressed by the product design measure.This measure ensures that installation by the user is clear and simplified, thereby effectively controlling the risk.
|