|
According to the complaint, on (b)(6) 2025 samples were measured on the abl90 flex analyzer (serial number: (b)(6) and showed a pco2 result of 10.7mmhg.This was inconsistent with the patient's condition.The user did not make any adjustments to the ventilator parameter of the patient, based on this measurement.Radiometer engineer was contacted, and after qc and inspection of the analyzer, the sensor cassette (lot number: r0287) was replaced.A repeat measurement was conducted and a pco2 result of 40.7mmhg was obtained.Based on this, the customer had reported the 10.7mmhg as false low.
|