| Model Number |
G6626525 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis #(b)(4):part # g6626525; lot # h5849260 visual and optical inspection revealed the implant was returned damaged.The implant has cracked next to where the release/lock mechanism is.The implant is fragile and can overloaded.It appears the implants structural integrity was overloaded during the implantation procedure.This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from health care professional via manufacturing representative regarding a patient undergoing for cervical spondylosis.It was confirmed that fracture of implanted fusion device during operation.Explant completed.No patient symptoms or complications related to this event.No further complications were reported/anticipated due to this event.Additional information was received from the rep that, cervical spondylosis was the pre op diagnosis for this event.
|
| |
|
Search Alerts/Recalls
|