SYNTHES GMBH OSCILLATING SAW ATCH, LARGE, WITH KEY; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
532.026 |
| Medical Device Problem Codes |
Device Slipped (1584); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a tibial plateau leveling osteotomy (tplo) surgical procedure it was observed that the tightening 'chuck' on the saw attachment device comes loose during the procedure.It was noted that the blade was secure, just the attachment on the handpiece was loose.It was further observed that the handpiece device had very slow reverse speed.It was not reported if there was delay in the procedure due to the event.It was not reported if there was a spare device available for use.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.D10.Concomitant med products and therapy dates: saw blade device, handpiece device ((b)(6) 2025).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: this device was returned for evaluation.A visual and functional assessment was performed which determined that the device passed all pre-repair diagnostic assessments, and no failures were identified.Therefore, the reported condition was not confirmed, and an assignable root cause was not determined.
|
| |
|
Search Alerts/Recalls
|
|
|