• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MAR NEUT 32IDX56OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MAR NEUT 32IDX56OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121932056
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 02/28/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D6a: implantation day was unknown date in 2014.H3, h6 investigation summary : the product was not returned to j&j medtech orthopaedics, however photos were provided for review.The photo investigation revealed that the pinn mar neut 32idx56od has slightly scratches at the rim most likely caused by the explantation process, however these scratches do not represent a device non conformance.Additionally at the evidence provided it can bee seen only the pinn mar neut 32idx56od assembled wit the pinnacle sector ii cup 56mm.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.A manufacturing records evaluation (mre) was not performed since the lot number was not able to be retrieved for the device due to the scratches.The overall complaint was not confirmed as the observed condition of the pinn mar neut 32idx56od would have not contributed to the reported adverse event.¿ based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A manufacturing records evaluation (mre) was not performed since the lot number was not able to be retrieved for the device due to the scratches.
 
Event or Problem Description
It was reported that there was a revision of a c-stem pinnacle.The reason for revision was because the stem was loose, and the hip was unstable.No allegation was made by the surgeon in relation to the loose stem.The cement/stem interface was loose.Cement bone/bone interface was sound.The surgeon noted that it was a very difficult primary.It was done by a different surgeon and they struggled to get a sz 1ho c-stem in.The x-ray showed that due to the anterior-lateral approach, the stem was put in varus.The hip was unstable because at the primary surgery, the cup was placed at the wrong version.The cup was explanted.It was extremely well fixed.The liner was explanted with the shell, but the lot number is unreadable as the liner was damaged when explanted.Reclaim bimentum was implanted.Affected side: right side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 32IDX56OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21687923
Report Number1818910-2025-04533
Device Sequence Number18462167
Product Code LPH
UDI-Device Identifier10603295013921
UDI-Public(01)10603295013921
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K001534
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number121932056
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2025
Initial Report FDA Received Date03/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-