• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA HP UNI INS SZ2 7MM RM/LL; EARLY INTERVENTION : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA HP UNI INS SZ2 7MM RM/LL; EARLY INTERVENTION : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 102454207
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the surgery, surgeon could not lock the insert.Another device was used to complete the surgery.There were no adverse consequences to the patient.There was no delay to surgery.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees hat the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: could not lock the insert.It was reported that during the surgery, could not lock the insert.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device revealed slight deformations and marks on the surface that are consistent with the implantation attempts.A manufacturing record evaluation was performed for the finished device [102454207 / m5748u] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified.The sigma ''high performance partial knee' unicondylar surgical technique (dsus/jrc/1114/0582 rev.3), provide guidance to for a correct final implantation steps to be followed.Following these steps will successfully locking the insert into the base as intended, preventing the uni insert to be damaged.The observed damaged condition suggests an improper alignment when trying to introduce the uni insert through the tibial base locking edge.This would contribute to the difficulty/inability to mate the insert with the tibial component following multiple insertion attempts.A functional test was not performed since mating device was not returned for evaluation.However, due to the visual damage observed, the difficulty/inability to assemble mating devices can be confirmed.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the sigma hp uni ins sz2 7mm rm/ll would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device [102454207 / m5748u] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified.Device history review: a manufacturing record evaluation was performed for the finished device [102454207 / m5748u] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified.H11 additional narrative: corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI INS SZ2 7MM RM/LL
Common Device Name
EARLY INTERVENTION : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21688475
Report Number1818910-2025-04540
Device Sequence Number9059312
Product Code NPJ
UDI-Device Identifier10603295002789
UDI-Public(01)10603295002789(17)290228(10)M5748U
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K070267
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number102454207
Device Lot NumberM5748U
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2025
Supplement Date Received by Manufacturer04/02/2025
04/11/2025
Initial Report FDA Received Date03/25/2025
Supplement Report FDA Received Date04/03/2025
04/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-