H6: code b20: the device remains implanted and, therefore, was not available for direct analysis by gore.H6: code c19: a review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.As the device was not accessible, the product history review (phr) review was the extend of the investigation.No device problem was found per review of these records.It should be noted that, per the gore® tag® thoracic branch endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to, endoleak.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 cfr part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
On (b)(6) 2025, patient underwent an endovascular procedure utilizing gore® tag® thoracic branch endoprosthesis.On 2/26/2025, fsa was notified an endoleak was observed from the side branch component in the left subclavian artery.It was also reported the side branch was not extended far enough distally during the initial implant procedure.The physician plans to bring the patient back for reintervention surgery on (b)(6) 2025.Physician plans to extend distally into the left subclavian artery to gain seal.No further patient information is available.
|