• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG FIBER OPTIC LIGHT CABLE, 250 CM, Ø4.8 MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG FIBER OPTIC LIGHT CABLE, 250 CM, Ø4.8 MM Back to Search Results
Model Number 495NCS
Medical Device Problem Code Device Handling Problem (3265)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 01/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: in the corresponding ifu fiber optic light cable, fluid light cables dsx551_en_v2.1_08-2024_ifu_ce-mdr, karl storz make clear reference to this under points 3.3 hot components and 3.4 high light intensity.On the optics-side connection, a melted adhesion (plastic) is visible on the front of the optical fibers which presumably originates from cover cloth contact.At the device-side connection, individual optical fibers are broken on the surface in certain areas of the device broken optical fiber are visible.This damage is due to normal use and the age of the light cable (january 2015) and was not caused by a manufacturing or production defect.The event is filed under internal karl storz complaint id: (b)(4).
 
Event or Problem Description
It was reported that the light source was left on top of the drapes and the patient was burnt on the face.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBER OPTIC LIGHT CABLE, 250 CM, Ø4.8 MM
Common Device Name
FIBER OPTIC LIGHT CABLE, 250 CM, Ø4.8 MM
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
christiane klaiber
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
MDR Report Key21689182
Report Number9610617-2025-00490
Device Sequence Number18462063
Product Code FCW
UDI-Device Identifier04048551001141
UDI-Public04048551001141
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number495NCS
Device Catalogue Number495NCS
Device Lot NumberNU02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Initial Report FDA Received Date03/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-