| Catalog Number |
67P040 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the trach would not inflate.Patient involvement is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event is unknown; no information has been provided to date.H3: device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: updated.H3: the device history record of reported lot number was reviewed and it was confirmed that no non-conformities were reported during production.The quality inspection forms were reviewed, and it was observed that no defects were found, and the lot was released.Complaint for the failure mode could not be confirmed by investigation as no samples nor pictures was provided by the customer.Without the return of the samples a comprehensive failure investigation cannot be performed and a probable cause cannot be determined.
|
| |
|
Search Alerts/Recalls
|