Model Number M00573060
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a jagtome rx 44 was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2025.During the procedure, the cutting wire of the jagtome rx 44 broke after energizing the third time.The device was approximately tensed for about one third.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with the original device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagtome rx 44 was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2025.During the procedure, the cutting wire of the jagtome rx 44 broke after energizing the third time.The device was approximately tensed for about one third.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with the original device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: (investigation result).The returned jagtome rx 44 was analyzed, and a visual and microscopic evaluation noted that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Based on the condition of the device, the wire break could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Jagtome RX 44
Common Device Name
Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela
CS  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key21691953
Report Number3005099803-2025-01227
Device Sequence Number18461958
Product Code KNS
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00573060
Device Catalogue NumberM00573060
Device Lot Number0035279803
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/10/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2025
Supplement Date Received by Manufacturer04/30/2025
07/03/2026
Initial Report FDA Received Date03/25/2025
Supplement Report FDA Received Date05/23/2025
07/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1