Model Number IN761S
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/24/2024
Type of Reportable Event Malfunction
Event or Problem Description
The device was returned to a third-party service center for evaluation in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The device was not in patient use.There was no report of serious patient harm or injury.During the evaluation of the device, the complaint was confirmed.The third-party service center visually inspected internal part of the device and found evidence of visible foam particles or degradation.Secondary findings include extreme dirt and moisture.In addition, the devices error log was downloaded, and one error code was present when reviewed.Lastly, device was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP AUTO BIFLEX
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 
4125423300
MDR Report Key21692286
Report Number2518422-2025-103400
Device Sequence Number8667835
Product Code BZD
UDI-Device Identifier00606959031531
UDI-Public(01)00606959031531
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIN761S
Device Catalogue NumberIN761S
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/24/2024
Initial Report FDA Received Date03/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/09/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1