MAUDE Adverse Event Report: SUNMED HOLDINGS LLC AIRFLOW; AIRFLOW PEDIATRIC RESUSCITATOR WITH INTEGRATED MANOMETER, B/V FILTER, 40CM POP-O

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SUNMED HOLDINGS LLC AIRFLOW; AIRFLOW PEDIATRIC RESUSCITATOR WITH INTEGRATED MANOMETER, B/V FILTER, 40CM POP-O Back to Search Results
Model Number AF2142MBS-45E
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
It has been reported that the airflow pediatric resuscitator mask being oval-shaped, making it difficult to connect to the bvm, is a reportable incident.This issue occurs during emergency situations, leading to delays in oxygen delivery due to the misshapen connection.Such delays could result in serious harm or injury if the patient does not receive oxygen quickly enough.
 
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 25 mar 2025 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.
 
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 25 mar 2025 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.The complaint of "the bvm mask hub is shaped like an oval" regarding part af2142mbs-45e was confirmed.The root cause was determined to originate from the supplier and was not caught during incoming inspection.A risk assessment was performed, and the ultimate risk was determined to be medium which does not require escalation to the capa review board.There have been 0 other complaints regarding the same part and a similar issue within the 24 months preceding this reported event.A resolution letter was sent to the customer.Complaints will continue to be monitored for possible trends.
 
Event or Problem Description
It has been reported that the airflow pediatric resuscitator mask being oval-shaped, making it difficult to connect to the bvm, is a reportable incident.This issue occurs during emergency situations, leading to delays in oxygen delivery due to the misshapen connection.Such delays could result in serious harm or injury if the patient does not receive oxygen quickly enough.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRFLOW
Common Device Name
AIRFLOW PEDIATRIC RESUSCITATOR WITH INTEGRATED MANOMETER, B/V FILTER, 40CM POP-O
Manufacturer (Section D)
SUNMED HOLDINGS LLC
2710 northridge dr nw suite a
grand rapids MI 49544
Manufacturer (Section G)
SUNMED AP, LTD.
6-9 zhufeng rd., qiannan indus
qianwu town, doumen district
zhuhai guangdong, china 51917 0
CH   519170
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key21692355
Report Number1314417-2025-00014
Device Sequence Number9419817
Product Code BTM
UDI-Device Identifier10889483565643
UDI-Public(01)10889483565643(10)490870
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAF2142MBS-45E
Device Catalogue NumberN/A
Device Lot Number490870
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2025
Supplement Date Received by Manufacturer02/25/2025
Initial Report FDA Received Date03/25/2025
Supplement Report FDA Received Date06/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-