TORNIER INC TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM STERILE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number DWJ718
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
Health Effect - Clinical Codes Pain (1994); Joint Dislocation (2374); Inadequate Osseointegration (2646)
Date of Event 02/24/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Patient underwent primary operation in (b)(6) 2023.Had increasing pain in operated shoulder and sought medical advice with non-primary surgeon.X-rays and ct scans showed baseplate was not fixed to scapular and was sitting within the joint.Also showed significant bone loss to remaining glenoid.Decision made to revise implant.Original implants removed.Bone loss too significant to do a single stage revision.Decision made to bone graft and let heal to be revised at a later date.Primary fixation of implant not achieved by primary surgeon.Loose implant caused significant bone loss in glenoid.Revision surgery required (multiple stages).
 
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Additional Manufacturer Narrative
Correction - please refer h6 device code.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
Patient underwent primary operation in (b)(6) 2023.Had increasing pain in operated shoulder and sought medical advice with non-primary surgeon.X-rays and ct scans showed baseplate was not fixed to scapular and was sitting within the joint.Also showed significant bone loss to remaining glenoid.Decision made to revise implant.Original implants removed.Bone loss too significant to do a single stage revision.Decision made to bone graft and let heal to be revised at a later date.Primary fixation of implant not achieved by primary surgeon.Loose implant caused significant bone loss in glenoid.Revision surgery required (multiple stages).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM STERILE
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
cécile lefeuvre
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key21695870
Report Number0001649390-2025-00168
Device Sequence Number8668035
Product Code KWS
UDI-Device Identifier00846832091225
UDI-Public(01)00846832091225(11)230627(17)280627(21)AF8767048
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K161742
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberDWJ718
Device Lot NumberAF8767048
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/24/2025
Supplement Date Received by Manufacturer04/22/2025
Initial Report FDA Received Date03/26/2025
Supplement Report FDA Received Date05/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-