• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OBSIDIO; VASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OBSIDIO; VASCULAR EMBOLIZATION DEVICE Back to Search Results
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Abdominal Pain (1685); Ischemia (1942); Necrosis (1971); Sepsis (2067); Multiple Organ Dysfunction Syndrome (3261); Embolism/Embolus (4438); Pancreatitis (4481)
Date of Event 02/09/2024
Type of Reportable Event Death
Additional Manufacturer Narrative
B3: date of event: the procedure date was between 9th and 12th february 2024.Jason s.Kim, md, r.Peter lokken, md, mph, alexander lam, md, neslihan kayraklioglu, md, phd, k.Pallav kolli, md."gastroduodenal artery embolization with shear-thinning conformable embolic resulting in pancreatic and duodenal necrosis".Journal of vascular and interventional radiology, 1- 17.
 
Event or Problem Description
It was reported via journal article, that migration occurred resulting in patient complications and death.Obsidio was selected for an embolization procedure of the posterior superior pancreaticoduodenal artery (pspda).The patient was admitted for hypoxemic respiratory failure and sweets syndrome, and the hospital course was complicated by melena.The patients hemoglobin declined overnight from 7.1 to 5.8 g/dl with recurrent melena and hematemesis requiring a blood transfusion.A computerized tomography (ct) angiogram demonstrated active extravasation into the duodenum.Angiography of the pspda with a 2.8 french non-boston scientific microcatheter demonstrated a pseudoaneurysm of a small branch where super selective catheterization was impossible.Embolization of the pspda using obsidio was performed, with the goal of occluding branches beyond the reach of the microcatheter.The aliquot technique was utilized to preserve the microcatheter for later use.While applying a slow injection force, 0.2 ml aliquots of obsidio was flushed with a 3 ml syringe and iohexol, which was surmised to better resemble the viscosity of obsidio compared to normal saline.The embolic was administered until opacification of the pseudoaneurysm was achieved, delivering a total of 0.4 ml.Fluoroscopic images demonstrated the embolic deposited in the pseudoaneurysm and distal branches measuring 0.2 to 0.3 mm in caliber, while forming a cast of the parent vessel.Angiography of the gastroduodenal artery (gda) demonstrated continued extravasation from a separate branch.Obsidio was therefore administered with a more forceful injection, with the intention of achieving gda occlusion without additional distal embolization, using 0.2 ml aliquots for a total of 0.6 ml.Although distal penetration with subsequent flushing was not directly observed, the embolic penetrated into the right gastroepiploic artery and other terminal branches measuring 0.4 mm; no reflux was appreciated.The patients post procedure course was complicated by abdominal pain and white blood cell count of 52,000/microliters, lipase of 240 u/l, and lactate dehydrogenase of 601 u/l.A ct of the abdomen/pelvis obtained on post procedure day 3, showed regional nonenhancement of the duodenum and pancreatic head.The patient was emergently taken for exploratory laparotomy where complete necrosis of the second portion of the duodenum and severe pancreatitis of the pancreatic head were observed.The necrotic duodenum was resected.The patient remained in critical condition with the postoperative course complicated by sepsis and multi-organ failure.The patient was transitioned to comfort care and succumbed 18 days after embolization.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OBSIDIO
Common Device Name
VASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
OBSIDIO, INC.
2 scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key21696264
Report Number2124215-2025-17223
Device Sequence Number9282269
Product Code KRD
UDI-Device Identifier00191506039332
UDI-Public00191506039332
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K213385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2025
Initial Report FDA Received Date03/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER: 2.8F PROGREAT TERUMO
Outcome Attributed to Adverse Event Death; Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-