| Model Number |
V273 |
| Medical Device Problem Codes |
Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439); Defective Device (2588)
|
| Health Effect - Clinical Codes |
Heart Block (4444); Insufficient Information (4580)
|
| Date of Event |
11/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Dimensional analysis of the header was completed.Each port measured as expected.Rv pace sense test was performed which passed and verified that a pacing signal was present.Interrogation of the device confirmed it was operating in safety mode and that bradycardia therapy remained available.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption (during telemetry or other higher power device operations) and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited oversensing on the right ventricular (rv) lead channel which led to pacing inhibition, which lasted less than two seconds.No additional adverse patient effects were reported.This device was explanted due to normal battery depletion.
|
| |
|
Search Alerts/Recalls
|