|
|
| Catalog Number |
ECR60W |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 03/26/2025.D4: batch#: unk.Additional information was requested, and the following was obtained: were the reloads (ecr60w) used with the device psee60a? yes.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device ecr60w with lot number: 519c14; and no non-conformance's were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic lobectomy procedure, it was observed that the staples malformed after firing the device.Changed to another one to continue the procedure but the same problem happened again.Another like device was used to complete the procedure.There was no patient consequence nor surgical delay reported.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 04/22/2025.D4: batch #519c14.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one psee60a device was returned with no apparent damage and with two ecr60w reloads present.The reloads were received unfired.The manual override door out of position; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, it was tested for functionality in the straight position with a returned reloads and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.The event described could not be confirmed as the device performed without any difficulties noted.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: ensure battery pack is fully inserted into the device.An audible click will be heard when the battery pack is fully inserted (the instrument must be used within 12 hours of inserting the battery pack, the battery pack contains a built-in battery drain).Insert the reload by sliding it against the bottom of the cartridge jaw until the cartridge alignment tab stops in the reload alignment slot.Snap the reload securely in place.Remove the staple retaining cap and discard.Close the jaws of the instrument by squeezing the closing trigger until it locks in place.An audible click indicates that the closing trigger and the jaws are locked.When the jaws of the instrument are closed, the red firing trigger lock and firing trigger are exposed.Pull back the red firing-trigger lock to enable the firing trigger to be pulled.Fire the instrument by pulling the firing trigger; the motor will activate audibly.Continue to depress the trigger until the motor stops.To complete the firing sequence, release the firing trigger to activate the motor and automatically return the knife to home position where the motor will stop.In this position, the instrument is locked out until the jaws are opened and re-closed.A manufacturing record evaluation was performed for the finished device ecr60w with lot number 519c14; and no non-conformances were identified.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|