| Model Number |
WA4KS400 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the rigid scope adapter came off in the light cord, during inspection, prior to patient in room for an unspecified procedure.There were no reports of patient harm.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
"this report is being supplemented to provide a correction to a previously submitted report.The event "adapter came off in the light cord" was reported in error, and it would be unlikely to cause or contribute to a death or a serious injury if it were to recur.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|