• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 104; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 104; GENERATOR Back to Search Results
Model Number GENERATOR, MODEL 104 HP
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 02/25/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had been experiencing pain and underwent a replacement.While the physician initially suspected infection, after the battery was removed, no infection was seen.The suspect device has not been received to date.No additional relevant information has been received to date.
 
Event or Problem Description
The suspect device was received into product analysis.Product analysis has not yet been completed.
 
Event or Problem Description
Product analysis was approved for the generator.An interrogation and a system diagnostic test were performed.The device output signal was monitored for more than 24-hrs and a comprehensive automated electrical evaluation (shows an ifi=no condition) was performed.No anomalies were seen, and the device performed according to functional specifications.There were no performance, or any other type of adverse conditions found with the pulse generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 104
Common Device Name
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics boulevard
suite 600
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics boulevard
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key21699127
Report Number1644487-2025-10007
Device Sequence Number18450245
Product Code LYJ
UDI-Device Identifier05425025750047
UDI-Public05425025750047
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/02/2017
Device Model NumberGENERATOR, MODEL 104 HP
Device Lot Number203361
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2025
Supplement Date Received by Manufacturer03/27/2025
06/04/2025
Initial Report FDA Received Date03/26/2025
Supplement Report FDA Received Date04/24/2025
06/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient SexMale
-
-