• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE; DQX WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Model Number G30506
Medical Device Problem Codes Flaked (1246); Chemical Problem (2893)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
As reported, prior to use for an unspecified diagnostic procedure, the "coating" came off a roadrunner uniglide hydrophilic wire guide.Per the reporter, a milky fluid formed from the tip of another manufacturer¿s pigtail catheter upon introduction of the wire into the catheter during preparation of the device.The device did not make patient contact.The patient did not require any additional procedures or experience any adverse effects due to this event.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1: phone= (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, prior to use for an unspecified diagnostic procedure, the "coating" came off a roadrunner uniglide hydrophilic wire guide.Per the reporter, a milky fluid formed from the tip of another manufacturer¿s pigtail catheter upon introduction of the wire into the catheter during preparation of the device.The device did not make patient contact.The patient did not require any additional procedures or experience any adverse effects due to this event.Corrected information: e1, h6 (annexes a & g).Investigation evaluation: reviews of the complaint history, device history record (dhr), manufacturing instructions, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformances on the final or sub-assembly lots.A review of complaint history found no additional complaints for this lot number.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr and dhr suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that a component failure contributed to this event.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key21699670
Report Number1820334-2025-00292
Device Sequence Number12927572
Product Code DQX
UDI-Device Identifier00827002305069
UDI-Public(01)00827002305069(17)270530(10)16082545
Combination Product (Y/N)N
PMA/510(K) Number
K130766
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG30506
Device Catalogue NumberHPWAS-35-260
Device Lot Number16082545
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2025
Supplement Date Received by Manufacturer06/20/2025
Initial Report FDA Received Date03/26/2025
Supplement Report FDA Received Date06/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORDIS PIGTAIL CATHETER.
Patient SexMale
-
-