• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SHPX535
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 11/26/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Photo evaluation summary: upon visual evaluation of the image provided in the complaint, the implant was observed ruptured.As the product involved in this complaint was discarded, the device could not be analyzed according to our procedures.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Mentor received an updated date of event and a confirmed serial number.Relevant fields have been updated.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient implanted with a 535 cc mentor memorygel xtra breast implant experienced a rupture on the left side post-operatively.As a result, the patient underwent explantation on (b)(6) 2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL XTRA BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key21699914
Report Number1645337-2025-03227
Device Sequence Number7910553
Product Code FTR
UDI-Device Identifier00081317025313
UDI-Public(01)00081317025313(17)251108(10)9516446(21)9516446-020
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2025
Device Catalogue NumberSHPX535
Device Lot Number9516446
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/05/2025
Supplement Date Received by Manufacturer04/04/2025
04/24/2025
Initial Report FDA Received Date03/26/2025
Supplement Report FDA Received Date04/22/2025
04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
Patient SexFemale
-
-