| Model Number |
N/A |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported there was concern for fracturing of internal stylet on insertion.Another picc had to be placed.No adverse event.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and it will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of broken stylet was confirmed.The product returned for evaluation was one 4fr dl powerpicc catheter w/ 3cg stylet.The stylet was evaluated and confirmed to be broken.The following observations were made which were consistent with this failure type: ¿ microscopic examination revealed localized delamination of the stylet tubing at the damage site(s), indicative of tubing kinking/folding ¿ microscopic examination revealed deformation of the stylet tubing at magnet edge(s) measurements of the stylet tubing wall thickness were taken and confirmed to be within specification.Although the break in the wire could be confirmed, the root cause could not be determined through examination of the returned sample.The observed features were indicative of circumstances which included stylet kinking, likely during the insertion process.It appeared that additional unidentified factors may have also contributed.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|
Search Alerts/Recalls
|