• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number 1109596
Medical Device Problem Codes Degraded (1153); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The foam degradation is covered under z-1958-2021.The ventilator inoperative is covered under z-1813-2024.
 
Event or Problem Description
A bipap a40 device was returned to a third party service center for servicing related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no reported patient involvement.During the evaluation of the device at third-party service center, visible foam particles were found, and the foam insulation was replaced to resolve the issue.Additionally, two ventilator inoperative error codes were found in the device's error log relating to a flow sensor communication as well as a pressure sensor communication error.The main printed circuit assembly (pca) board was replaced to resolve the issues.Additionally, a low circuit leak error code found in the devices error log.It is also noted that extreme dirt and moisture were found during the evaluation.The device software was upgraded, and the device passed all final testing.No further investigation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key21700143
Report Number2518422-2025-103438
Device Sequence Number18120246
Product Code MNT
UDI-Device Identifier00606959042742
UDI-Public(01)00606959042742
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1109596
Device Catalogue Number1109596
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2024
Initial Report FDA Received Date03/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1958-2021; Z-1813-2024
Patient Sequence Number1
Patient SexUnknown
-
-