• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE MECHANICAL LITHOTRIPTOR V

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE MECHANICAL LITHOTRIPTOR V Back to Search Results
Model Number BML-V242QR-30
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during endoscopic retrograde cholangiopancreatography (ercp) procedure, the wire of the single use mechanical lithotriptor broke while the basket wire was inside the patient with a stone in it.Troubleshooting was attempted with a use of a different mechanical lithotriptor but the wire broke again.The physician used other means to remove the basket wire by cutting two of the four basket wires using an argon plasma coagulation and then was able to release the stone.This resulted to the delay of the procedure for at least an hour.The procedure was then completed with a similar device.The physician stated there were no complications but the stone was unable to be crushed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE MECHANICAL LITHOTRIPTOR V
Common Device Name
SINGLE USE MECHANICAL LITHOTRIPTOR
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
MDR Report Key21700226
Report Number2429304-2025-00110
Device Sequence Number8635973
Product Code LQC
UDI-Device Identifier04953170218392
UDI-Public04953170218392
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/26/2025
Report Date (Section F) 03/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBML-V242QR-30
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date02/24/2025
Event Location Hospital
Date Report to Manufacturer02/24/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/26/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BML-110A-1 (MECHANICAL LITHOTRIPTOR)
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Date Report Sent to FDA03/26/2025
-
-