| Model Number |
H7493960830400 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a balloon rupture occurred.The target lesion was located in the proximal right coronary artery.The 4.00 x 30mm agent dcb was introduced to treat the target lesion but the balloon material ruptured.The procedure was completed with a balloon angioplasty catheter resulting in 25% residual stenosis and timi 3 flow.
|
| |
|
Event or Problem Description
|
|
It was reported that a balloon rupture occurred.The target lesion was located in the proximal right coronary artery.The 4.00 x 30mm agent dcb was introduced to treat the target lesion but the balloon material ruptured.The procedure was completed with a balloon angioplasty catheter resulting in 25% residual stenosis and timi 3 flow.It was additionally reported that the target lesion was 50% stenosed, and the agent dcb was inflated to 14 atm for 25 to 30 seconds when it ruptured.
|
| |
|