|
Section d5: operator of device corrected manufacturer's investigation conclusion: the report of a blood leak from the oxygenator was unable to be confirmed as no product or images were available for evaluation.A specific root cause for the reported event could not be conclusively determined through this evaluation; however, eurosets determined that the reported leak can be related to fiber rupture.The patient was on pediatric eurosets oxygenator support for two days when the oxygenator began leaking blood.Whole red blood was reported in the lower portion of the oxygenator, within the gas outlet.It was reported that there was no increase in pressure noted and no decrease in gas transfer.The patient was stable on support when the decision was made to exchange the oxygenator.Pump speed was 2850 revolutions per minute, blood flow was 1.28 liters per minute, inlet pressure was 200 millimeters of mercury (mmhg), and outlet pressure was 176 mmhg.The patient¿s activated clotting time/partial thromboplastin time was 70.8.The production documentation for the eurosets pediatric oxygenator, lot number 10453500, was reviewed by the external manufacturer (eurosets) and showed that all tests from the production process were compliant with the technical specifications.Devices are required to pass manufacturing inspections and specifications prior to release and no abnormalities were documented which would cause or contribute to the reported occurrence.This device passed all required testing.Eurosets confirmed that 100% of the oxygenators produced are tested to detect eventual leakages using a pressure of 150 kpa (kilopascals) which is 1.5 times the maximum blood pathway pressure indicated on the ifu (instructions for use).Devices that exhibit leaks are discarded prior to distribution.Eurosets determined that the leak reported during use can be related to a fiber rupture after eurosets manufacturing and test phases.Eurosets communicated that their production process and controls will continue to be improved to further reduce the occurrence rate of these events, which will also be monitored for adverse trends.The production documentation for the eurosets oxygenator, lot number 10453500, was reviewed by the external manufacturer (eurosets) and showed that all tests made in the production process were compliant with the technical specifications.The eurosets amg pmp pediatric instructions for use (ifu), is currently available.Under ¿risks and side effects,¿ the ifu lists possible risks and side effects, including blood leakage caused by mechanical failure (e.G.Loss of mechanical integrity).These are potential side effects of all extracorporeal blood circulation systems.Under the list of warnings, the ifu warns that during use, a spare oxygenator must always be available and also warns that the device has to be carefully and continuously checked by qualified healthcare professionals throughout the procedure.The ifu warns to carefully check the device seal during priming and operation.Any loss may result in contamination of the environment or the operator, loss of blood, air embolism.If you notice leakage during priming or operation, replace the defective device following good perfusion practices.Also, under the section titled ¿set up¿, the ifu warns to ¿carry out a visual inspection and carefully check the device before use.Transport and/or storage conditions other than those prescribed may have caused damage to the device.¿ under the section titled ¿during bypass¿, the ifu instructs that during the entire procedure, according to good prefusion practices, monitor the integrity of the system for leaks absence.In this section, the ifu warns that ¿during cpb (cardiopulmonary bypass) check the oxygenator integrity, a small breach can also result in continuous blood loss, bacterial risk, viral infection or pyrogen reaction.A large breach can result in rapid blood loss leading to shock and death.If leakage occurs replace the oxygenator with a new one.¿ under the section titled ¿oxygenator replacement¿, this document states that a spare oxygenator must always be available during perfusion.After 6 hours of use with blood or if particular situations occur, which may lead the person responsible for perfusion to determine the safety of the patient may be compromised (insufficient oxygenator performance, leaks, abnormal blood parameters, etc.), follow the procedure outlined in the ifu for oxygenator replacement.No further information was provided.The manufacturer is closing the file on this event.
|