COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number G48295
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the ngage nitinol stone extractor¿s lower part of the basket was noted to be broken during a lithotripsy procedure for removal of a kidney stone.The procedure was completed by using another same-like device.A customer provided picture appears to show separation of the basket from the basket sheath.As section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
A5: ethnicity= chinese.G4: pma/510(k) #: exempt.H3: device has not yet been returned to manufacturer; however, our investigation is ongoing, and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
This report is being sent to indicate the complaint event is not reportable under fda 21 cfr part 803.An mdr guidance for manufacturers issued in 1997 stated that once a malfunction has caused or contributed to a death or serious injury, a presumption that the malfunction is likely to cause or contribute to a death or serious injury has been established.This presumption will continue until either the malfunction has caused or contributed to no further deaths or serious injuries for two years, or the manufacturer can show through valid data that the likelihood of another death or serious injury as a result of the malfunction is remote.A validated and detailed search of cook¿s complaint handling software revealed that there have been no deaths or serious injuries per 21 cfr part 803.3, due to the ngage nitinol stone extractor basket wires tore from the basket tubing, from 01jan2022 through 11aug2025.Therefore, cook will cease malfunction reporting for events where the ngage nitinol stone extractor basket wires tore from the basket tubing.The recurrence of a serious injury or death for the same malfunction will trigger the resumption of mandatory reporting, per 21 cfr part 803.50.As such, this event no longer meets the set criteria for a reportable event; no further reports regarding this event will be submitted.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Common Device Name
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key21700976
Report Number1820334-2025-00294
Device Sequence Number8593386
Product Code FFL
UDI-Device Identifier00827002482951
UDI-Public(01)00827002482951(17)260605(10)15473743
Combination Product (Y/N)N
Initial Reporter CountryCN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/05/2026
Device Model NumberG48295
Device Catalogue NumberNGE-017115-MB
Device Lot Number15473743
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2025
Supplement Date Received by Manufacturer07/30/2025
Initial Report FDA Received Date03/26/2025
Supplement Report FDA Received Date08/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BOSON DISPOSABLE FLEXIBLE URETEROSCOPE.
Patient SexMale
-
-