| Catalog Number |
168-390 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side exchange of textured devices due to product concern and capsular contracture baker grade iii.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h.6.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: a2, a4, a5, a6, d9, h3, h6.Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: - capsular contracture: unable to observe as it is not related to the manufacturing process.- anxiety-product/procedure: unable to observe as it is not related to the manufacturing process.As per the investigation procedure, creases and wear abrasion were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side exchange of textured devices due to product concern and capsular contracture baker grade iii.Device has been explanted.
|
| |
|