MAUDE Adverse Event Report: SYNTHES GMBH 10MM/125 DEG TI CANN TFNA 360MM/LEFT - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 10MM/125 DEG TI CANN TFNA 360MM/LEFT - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 04.037.027S
Medical Device Problem Codes Device-Device Incompatibility (2919); Difficult to Advance (2920)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on march 4, 2025, the patient underwent an unknown surgery with tfna.The surgeon tried to insert the end cap after nail insertion.No matter how many times the surgeon reinserted it, it never completely inserted into the nail hole.Imaging showed that it looked like it was floating about 1 mm.The hospital had two end caps and the surgeon tried the other one.However, the same event occurred.Therefore, the surgeon guessed that the nail had a problem, and the procedure was completed as it was.The locking mechanism was re-engaged with a torque screwdriver.There was no strange sound during insertion.The surgery was completed successfully without any surgical delay.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3, h6: manufacturing location: monument; manufacturing date: 13-august-2020; expiration date: 1-august-2030; part: 04.037.027s, 10mm/125 deg ti cann tfna 360mm/left ¿ sterile; lot: 65p5573.Production order traveler met all inspection acceptance criteria.Inspection sheet, in-process / inspect dimensional / final met all inspection acceptance criteria.Inspection sheet, tfna assembly inspection met all inspection acceptance criteria.Packaging label log (pll) was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Sterilization control number (scn) supplied by ethicon (abq) was reviewed and determined to be conforming.This lot met all dimensional, visual, sterilization and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM/125 DEG TI CANN TFNA 360MM/LEFT - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key21701376
Report Number8030965-2025-02811
Device Sequence Number7868246
Product Code HSB
UDI-Device Identifier10886982095855
UDI-Public(01)10886982095855(17)20300801(10)65P5573
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.037.027S
Device Lot Number65P5573
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/03/2025
Initial Report FDA Received Date03/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-