• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY SYSTEM 2; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY SYSTEM 2; PLATE Back to Search Results
Model Number SK14
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 02/27/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that after an initial bunion surgery on (b)(6) 2024, all hardware was removed and replaced during a revision surgery on (b)(6) 2025 due to non-union.No other patient outcomes were reported as a result of this event.No devices were returned for evaluation.The device history records were reviewed and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.Although several factors can contribute to the reported event, the most likely cause cannot be determined due to the limited information provided, and no device being returned.No deficiencies or malfunctions have been alleged/found with any tmc device.The company will supplement this mdr as necessary and appropriate.
 
Event or Problem Description
It was reported that after an initial bunion surgery on (b)(6) 2024, all hardware was removed and replaced during a revision surgery on (b)(6) 2025 due to non-union.No deficiencies or malfunctions were alleged/found with any tmc device, and no other patient outcomes were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPIPLASTY SYSTEM 2
Common Device Name
PLATE
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
renae ginter
100 palmetto park place
ponte vedra, FL 
9043735940
MDR Report Key21704528
Report Number3011623994-2025-00030
Device Sequence Number9941856
Product Code HRS
UDI-Device Identifier00853114006051
UDI-Public(01)00853114006051(17)290419(10)43133
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K220136
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSK14
Device Lot Number43133
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2025
Initial Report FDA Received Date03/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-