MAUDE Adverse Event Report: COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number G48294
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 - customer phone number: (b)(6).G4: pma/510(k) #: exempt h3: device evaluation anticipated, but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported the ngage nitinol stone extractor¿s basket wire was broken when the package was opened, prior to an unspecified procedure.The procedure was completed by using another same-like device.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.The device did not make patient contact.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Summary of event: it was reported the ngage nitinol stone extractor¿s basket wire was broken when the package was opened, prior to an unspecified procedure.The procedure was completed by using another same-like device.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.The device did not make patient contact.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions (mi), and quality control (qc) procedures, a visual inspection, and functional test of the returned device, were conducted during the investigation.One ngage nitinol stone extractor was returned in an open package with label.The basket wire is broken, there no signs of charging or deformity that could have caused the break.The handle will actuate basket.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that that the device was manufactured to specification and there is no evidence that nonconforming product exists in house or in the field.A review of the ifu provides the following information: suggested handling instructions for extractors and forceps: important: excessive force could damage device.Based on the information provided and the results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Common Device Name
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key21707657
Report Number1820334-2025-00298
Device Sequence Number11961287
Product Code FFL
UDI-Device Identifier00827002482944
UDI-Public(01)00827002482944(17)270701(10)16135836
Combination Product (Y/N)N
Initial Reporter CountryCN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG48294
Device Catalogue NumberNGE-017115
Device Lot Number16135836
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2025
Supplement Date Received by Manufacturer08/01/2025
Initial Report FDA Received Date03/27/2025
Supplement Report FDA Received Date08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-