MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3544800 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
05/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation review could be performed.Reason for device explant and/or reoperation: left rupture.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 72-year-old caucasian female patient underwent primary breast augmentation with 800cc mentor memorygel breast implant and experienced breast implant rupture on her left side postoperatively.As a result, the patient underwent replacement surgery on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
On april 8, 2025, the mentor failure analysis lab received two devices for evaluation, it cannot be determined which device is the suspect medical device for the reported rupture, since the two received products have the same volume engraved, both devices were found damaged per analysis findings.Hence, both device result are being submitted.Hence, this report is for the left side.On april 15, 2025, device evaluation was completed as follows: mentor conducted a visual inspection.Visual analysis of the returned device revealed that the siltex hpg, 800cc breast implant was found to be ruptured.As the authorization form for examination was not received, the product evaluation lab could not reach a conclusion regarding the specific nature of the ruptures.The evaluation was limited to non-destructive testing.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause the implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number (b)(4).
|
| |
|