| Model Number |
1475 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the pioneer convertible patient pack experienced damage, specifically with the velcro being frayed and peeling up.There were no patient symptoms or complications associated with this event.The patient pack was replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for analysis and investigation completion.Product event summary: the patient pack (lot no.0008550) was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.A review of the manufacturing documentation was not performed as the reported event and analysis are not related to a manufacturing or servicing issue.Failure analysis of the returned patient pack revealed a damaged velcro patch.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to wear and/or the handling of the pack.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|