| Model Number |
21-7310-24 |
| Medical Device Problem Codes |
Failure to Infuse (2340); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while the cassette was connected to the pump, the device exhibited an occlusion alarm.There was patient involvement, no patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation.A functional test was performed and the reported issue was not duplicated.The cause of the reported issue was unknown.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
|
| |
|