ABBOTT VASCULAR INC. COPILOT; ADAPTOR, STOPCOCK, MANIFOLD, FITTING, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 1003331
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.In this case, it is possible that the cap seal was not fully tightened, therefore resulting in the reported leak difficulties.However, the device was not returned for analysis; therefore, it cannot be confirmed.The investigation determined a conclusive cause for the reported difficulty cannot be determined.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a right coronary artery.The copilot bleedback control valve was noted to have constant air getting into syringe when drawing back on the syringe.The same device was used throughout the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COPILOT
Common Device Name
ADAPTOR, STOPCOCK, MANIFOLD, FITTING, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key21712408
Report Number2024168-2025-03162
Device Sequence Number20019901
Product Code DTL
UDI-Device Identifier08717648013997
UDI-Public(01)08717648013997(17)260630(10)60573590
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K991102
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model Number1003331
Device Catalogue Number1003331
Device Lot Number60573590
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2025
Initial Report FDA Received Date03/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-