Model Number CCWET4
Medical Device Problem Code Failure to Unfold or Unwrap (1669)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history record could not be reviewed because the reporting facility did not provide a valid lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during surgery the surgeon stated lens would not unfold in the cartridge and there was no patient contact.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information was provided in d.4.,h.3., h.6.And h.11.The account indicated the use of a qualified associated cartridge and handpiece with a non-qualified viscoelastic.The root cause for the reported complaint may be related to a failure to follow the ifu.The viscoelastic indicated is not qualified for the lens/company combination used.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu (instructions for use) instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd (ophthalmic viscosurgical device) filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol (intraocular lens) to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The product has not been received to evaluate.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON VIVITY TORIC EXTENDED VISION UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21717779
Report Number1119421-2025-00863
Device Sequence Number17271797
Product Code MJP
UDI-Device Identifier00380652455695
UDI-Public(01)00380652455695(17)270911(11)230913(10)15711161
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCCWET4
Device Catalogue NumberCCWET4.225
Device Lot Number15711161
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/27/2025
Supplement Date Received by Manufacturer04/30/2025
Initial Report FDA Received Date03/28/2025
Supplement Report FDA Received Date05/26/2025
Date Device Manufactured09/13/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV IOL, INJECTOR; EYEFILL, LOT 207011; MONARCH III IOL, CARTRIDGE D; TIER FORCEPS
Patient SexUnknown
-
-