• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSURE SP NOVUS LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSURE SP NOVUS LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 4592-53
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: 5076-58 lead; implant date: (b)(6) 2010.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the right atrial (ra) lead exhibited high impedance and noise.It was noted the lead had a possible fracture also.It was further noted the lead was reprogrammed.The lead was removed and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the atrial pacing lead was beyond the expected upper range.Analysis of the device memory indicated atrial pacing capture threshold was elevated.Analysis of the device memory indicated atrial oversensing.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SP NOVUS LEAD
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key21717948
Report Number2649622-2025-08501
Device Sequence Number17314071
Product Code DTB
UDI-Device Identifier00681490114974
UDI-Public00681490114974
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P830061
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/04/2011
Device Model Number4592-53
Device Catalogue Number4592-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2025
Supplement Date Received by Manufacturer03/31/2025
Initial Report FDA Received Date03/28/2025
Supplement Report FDA Received Date04/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
W3DR01.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-