Model Number CB11BA40D
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that both burs broke after a few seconds during the set up.There were no fragments left inside the patient and there was no delay in procedure as a result of the event.It was reported that there was no patient impact.The procedure was completed with the backup product.Upon follow up it was confirmed that there was no additional intervention was performed and the procedure was spine surgery.
 
Additional Manufacturer Narrative
H3: product analysis: this is well-known failure mode.The most likely cause of failure is variation in the manufacturing process that caused tang weld to fail after some use, leading to fatigue failure at the stem.This regulatory report is being submitted due to a retrospective review through capa: 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21718034
Report Number1625507-2025-01136
Device Sequence Number4514857
Product Code HBE
UDI-Device Identifier00643169766228
UDI-Public00643169766228
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K072315
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/20/2025
Device Model NumberCB11BA40D
Device Catalogue NumberCB11BA40D
Device Lot Number0222732145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2023
Initial Report FDA Received Date03/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-