• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION NX19 OXYGENATOR W/4-LITER RESERVOIR; LEFT-SIDE OUTLET PORT; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION NX19 OXYGENATOR W/4-LITER RESERVOIR; LEFT-SIDE OUTLET PORT; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 3CX*NX19RW
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, there was a plasma leak.No health consequences or impact.Product was not changed out.Procedure completed successfully.
 
Additional Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 28, 2025.Upon further investigation of the reported event, the following information is new and/or changed: g3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H6 (identification of evaluation codes 10, 3331, 4210, 4307).The returned sample was inspected upon receipt and noted to have red tinged fluid inside of the gas caps.The unit was decontaminated upon receipt as per protocol.During the decontamination process, foaming was observed in the gas out cap.It is possible that there are unknown factors that contributed to the plasma leakage.A definitive root cause was unable to be determined.The plasma leakage is believed to be a result of prolonged usage which led to the fiber pore surfaces becoming hydrophilic through absorption of elements circulating in the blood over time.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NX19 OXYGENATOR W/4-LITER RESERVOIR; LEFT-SIDE OUTLET PORT
Common Device Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
125 blue ball road
elkton, MD 21921
7346634145
MDR Report Key21718944
Report Number1124841-2025-00026
Device Sequence Number19779138
Product Code DTZ
UDI-Device Identifier00699753450875
UDI-Public(01)00699753450875(17)250930(10)CM30
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K180950
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model Number3CX*NX19RW
Device Catalogue NumberN/A
Device Lot NumberCM30
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2025
Supplement Date Received by Manufacturer04/29/2025
Initial Report FDA Received Date03/28/2025
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-