• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON PANOPTIX TRIFOCAL UV ABSORBING INTRAOCULAR LENS; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON PANOPTIX TRIFOCAL UV ABSORBING INTRAOCULAR LENS; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number CCWTT0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Type of Reportable Event Serious Injury
Event or Problem Description
A non health care professional reported that following an intraocular lens (iol) implant procedure, the patient had intolerable nighttime star-bursting/halos.Clinical reason being mentioned as dysphotopsia.The iol was explanted and exchanged for an unknown monofocal lens in the secondary implant procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Additional Manufacturer Narrative
Additional information has been provided in h.3., h.6., and h.11.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.The product has not been received to evaluate.Each lens is subjected to a 100 percent assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the multifocal company, 17.0 diopter (add power plus 2.2/3.2) lens was replaced with an unspecified, monofocal lens model.Due to the exchange of a multifocal lens to a monofocal lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.No additional information has been provided at this time.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to requests for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TRIFOCAL UV ABSORBING INTRAOCULAR LENS
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21719697
Report Number1119421-2025-00866
Device Sequence Number3783946
Product Code MFK
UDI-Device Identifier00380652413589
UDI-Public(01)00380652413589(17)250408(11)211115(10)15287999
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/08/2025
Device Model NumberCCWTT0
Device Catalogue NumberCCWTT0.170
Device Lot Number15287999
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2025
Supplement Date Received by Manufacturer05/29/2025
Initial Report FDA Received Date03/28/2025
Supplement Report FDA Received Date06/23/2025
Date Device Manufactured11/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-