MAUDE Adverse Event Report: MICROVENTION INC. AZUR-18 DETACHABLE HYDRO; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION INC. AZUR-18 DETACHABLE HYDRO; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 45-480405-L
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was stated to be available for return to the manufacturer for evaluation, but has not yet been returned.The alleged product issue could not be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with the use of the device.
 
Event or Problem Description
As reported in an interventional radiology, an azur coil was used during the procedure.The coil detached spontaneously from its delivery catheter into the microcatheter, without manipulation error.The device was removed, and a second coil was used.There were no clinical consequences for the patient.
 
Additional Manufacturer Narrative
Investigation conclusion: the investigation of the returned coil system found the implant stretched, damaged, and separated from the pusher with the monofilament exposed; however, no indications of activation using a detachment controller was found on the pusher heater coil.The exposed monofilament was found to have a tensile break shape at the tip, which is consistent with the device experiencing force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the implant damage, but the damage is consistent with the device experiencing forces exceeding the implant's yield strength.The microcatheter used in the procedure was not returned, so the investigation could not determine if it had caused or contributed to the reported complaint.
 
Event or Problem Description
No additional information was received, please see h6 & h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR-18 DETACHABLE HYDRO
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key21724168
Report Number2032493-2025-90024
Device Sequence Number11744512
Product Code KRD
UDI-Device Identifier00810170017896
UDI-Public(01)00810170017896(11)230426(17)280331(10)0000355379
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K090168
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 05/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number45-480405-L
Device Lot Number0000355379
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2025
Supplement Date Received by Manufacturer05/19/2025
Initial Report FDA Received Date03/28/2025
Supplement Report FDA Received Date05/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-