MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507600BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Breast Discomfort/Pain (4504)
|
| Date of Event |
08/20/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast pain, material rupture mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent breast augmentation revision with two 600cc mentor memorygel breast implants.Post-operatively, the patient developed right breast pain and was diagnosed, via mri, with possible bilateral breast implant ruptures.As a result, the patient underwent bilateral breast implant removal and replacement surgery on (b)(6) 2025.During the surgery, the patient was diagnosed with deformity but only the right breast implant was found to be ruptured.The left implant was removed intact.Subsequently, they were both replaced with the following: (left) 600cc mentor memorygel breast implant catalog: 3506001bc lot: 9997828 sn: (b)(6) and (right) 600cc mentor memorygel breast implant catalog: 3506001bc lot: 9997828 sn: (b)(6).This medwatch form is for the right breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On june 13, 2025, the mentor failure analysis lab received the device for evaluation.On june 20, 2025, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned device revealed that the gel mod-rnd 600cc breast implant was ruptured.In addition, shell abrasion was observed at the edges of the rupture.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Surgeons should instruct their patients to inform them if there is significant pain or if pain persists.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.
|
| |
|
Additional Manufacturer Narrative
|
|
On july 22, 2025, mentor received additional information that indicated that the patient experienced breast pain for six months prior to them getting an mri of their breast.The date of event was updated to its best estimate and has been populated under section b.Date of event.
|
| |
|