|
Result of investigation: a visual and functional test was carried out on the endoscope.The endoscope shows signs of wear and scratches.The steering mechanism rattles and scrapes.The vertebrae and the angle cover are mechanically damaged, causing the endoscope to leak.There is moisture in the housing.The sensor and the interface board are damaged by moisture.The "no image" error described by the customer was confirmed.This was caused by a mechanical impact that compressed and deformed the vertebrae and damaged the angle cover, causing it to leak.Moisture has entered the housing through the leaking angle cover.The ingress of moisture caused the sensor and the interface board to fail.For this reason, a user misuse error is to be assumed which led to the missing image.Endoscope bending section is heavily damaged, by excessive mechanical force, due to abnormal use.The above conducted investigations did not reveal a root cause related to the labelling, the device or its history.All production related quality checks were passed, and no non-conformities were identified in the devices manufacturing records.The labelling was found to contain adequate instructions and warnings.The complaint history did not reveal any pickup in similar complaints, no trend was identified.According to the ifu (96056039d, version v3.1_01-2020) functionality should always be checked before every use to avoid injuries and damages to the product.Damaged scopes must not be used.It is described in detail how the scope must be handles stored correctly to avoid damage.The event is filed under internal karl storz complaint id: (b)(4).
|