Model Number 328440
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported liquid inside of the barrel of his syringes.He stated that when he depressed the plunger rod, he noticed the liquid coming out.Consumer stated that the bag was sealed.Lot #: 4191017, catalog #: 328440, date of event: unknown, samples: available - sending mail kit.
 
Additional Manufacturer Narrative
Additional information was added to: b4, g6, h2, h3, h11.Correction to: h6 (investigation findings and investigation conclusions).Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue as silicone.Based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 0.3ML 31GA 8MM HALFUNIT 10BG 500
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
EMBECTA MEDICAL I LLC - HOLDREGE, NE
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key21726842
Report Number3024508819-2025-00157
Device Sequence Number11728019
Product Code FMF
UDI-Device Identifier00382908440035
UDI-Public00382908440035
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K212499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number328440
Device Lot Number4191017
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/28/2025
Supplement Report FDA Received Date04/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-