|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, d9, h2, h3, h4, h6 and h11.The device was returned to the regional service center for inspection and the reported failure was confirmed.In addition, the following reportable malfunction was identified during the device evaluation: dent on the outer tube and scratches at the distal end.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: regarding the customer¿s allegation, since the lens is cracked near adaptor end is cause traced to component failure.And for the newly identified malfunction mentioned above, the cause was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|